IFM PN2592 压力传感器

IFM

IFM PN2592 压力传感器

带显示屏的压力传感器

IFM PN2592 压力传感器 — datasheet、规格参数、订货号查询与原装正品采购

获取报价 / 询盘

立即询价 WhatsApp 询盘 邮件询价

产品亮点

产品概述

产品标识电子压力传感器
应用用于工业应用
系统接口螺纹连接 G 1/4 外螺纹 (DIN EN ISO 1179-2); 内螺纹:M5
工作电压DC18...30 V
电压类型DC
电气设计PNP/NPN
输出数量2
输出功能常开/常闭; (可设定参数); 模拟
输出信号开关信号; 模拟信号; IO-Link; (可配置)
模拟电流输出4...20; (可调整量程 1:5) mA
模拟电压输出0...10; (可调整量程 1:5) V
测量范围0...100 bar0...1450 psi0...10 MPa
接口M12 接插件
原材料不锈钢(1.4404 / 316L); PBT+PC-GF30; PBT-GF20; PC
材料(接液部件)不锈钢(1.4404 / 316L); Al2O3 (陶瓷); FKM
显示显示单位: 3 x LED 绿色 (bar, psi, MPa); 开关状态: 2 x LED 黄色; 测量值: 字母数字显示 红色/绿色 4位数字
环境温度-25...80 °C
介质温度-25...80 °C
通信接口IO-Link
产品易福门或同价值的
型号PN2592

技术规格

Product characteristics
Number of inputs and outputsNumber of digital outputs: 2; Number of analogue outputs: 1
Measuring range0...100 bar 0...1450 psi 0...10 MPa
Process connectionthreaded connection G 1/4 external thread (DIN EN ISO 1179-2); internal thread:M5
Application
Special featureGold-plated contacts
Measuring elementceramic-capacitive pressure measuring cell
Applicationfor industrial applications
MediaLiquids
Conditionally suitable foruse in gases at pressures > 25 bar only on request
Medium temperature[°C]-25...80
Min. burst pressure650 bar 9400 psi 65 MPa
Pressure rating300 bar 4350 psi 30 MPa
Vacuum resistance
-1000 mbar-0.1 MPa
Type of pressurerelative pressure
Electrical data
Operating voltage[V]18...30 DC; (to SELV/PELV)
Current consumption[mA]< 35
Min. insulation resistance[MΩ]100; (500 V DC)
Protection classIII
Reverse polarity protectionyes
Power-on delay time[s]0.3
Integrated watchdogyes
Inputs / outputs
Number of inputs and outputsNumber of digital outputs: 2; Number of analogue outputs: 1
Outputs
Total number of outputs2
Output signalswitching signal; analogue signal; IO-Link; (configurable)
Electrical designPNP/NPN
Number of digital outputs2
Output functionnormally open / normally closed; (parameterisable)
Max. voltage drop switching output DC[V]2
Permanent current rating of switching output DC[mA]250
Switching frequency DC[Hz]< 500
Number of analogue outputs1
Analogue current output[mA]4...20; (scalable 1:5)
Max. load[Ω]500
Analogue voltage output[V]0...10; (scalable 1:5)
Min. load resistance[Ω]2000
Short-circuit protectionyes
Type of short-circuit protectionpulsed
Overload protectionyes
Measuring/setting range
Measuring range0...100 bar 0...1450 psi 0...10 MPa
Analogue start point0...80 bar 0...1160 psi 0...8 MPa
Analogue end point20...100 bar 290...1450 psi 2...10 MPa
Factory setting / CMPT = 2
Set point SP0.6...100 bar 10...1450 psi 0.06...10 MPa
Reset point rP0.2...99.6 bar 4...1444 psi 0.02...9.96 MPa
Min. difference between SP and rP0.6 bar 6 psi 0.06 MPa
In steps of0.2 bar 2 psi 0.02 MPa
Status_B High Resolution / CMPT = 3
Set point SP0.6...100 bar 9...1450 psi 0.06...10 MPa
Reset point rP0.2...99.6 bar 3...1444 psi 0.02...9.96 MPa
Min. difference between SP and rP0.5 bar 6 psi 0.05 MPa
In steps of0.1 bar 1 psi 0.01 MPa
Accuracy / deviations
Switch point accuracy[% of the span]< ± 0,4; (Turn down 1:1)
Repeatability[% of the span]< ± 0,1; (with temperature fluctuations < 10 K; Turn down 1:1)
Characteristics deviation[% of the span]< ± 0,25 (BFSL) / < ± 0,5 (LS); (Turn down 1:1; BFSL = Best Fit Straight Line; LS = limit value setting)
Hysteresis deviation[% of the span]< ± 0,1; (Turn down 1:1)
Long-term stability[% of the span]< ± 0,05; (Turn down 1:1; per 6 months)
Temperature coefficient zero point[% of the span / 10 K]< ± 0,2; (-0...80 °C)
Temperature coefficient span[% of the span / 10 K]< ± 0,2; (-0...80 °C)
Notes on the accuracy / deviationswitch point accuracy, linearity error under DNV GL: < ± 1%: < ± 1%
Response times
Response time[ms]< 1.5
Delay time programmable dS, dr[s]0...50
Damping process value dAP[s]0...4
Damping for the analogue output dAA[s]0...4
Max. response time analogue output[ms]3
Software / programming
Parameter setting optionshysteresis / window; normally open / normally closed; switch-on/switch-off delay; Damping; Display unit; current/voltage output
Interfaces
Communication interfaceIO-Link
Transmission typeCOM2 (38,4 kBaud)
IO-Link revision1.1
SDCI standardIEC 61131-9
SIO modeyes
Required master port typeA; (when pin 2 not connected: B)
Supported DeviceIDs
Type of operationDeviceID
Factory setting / CMPT = 2461
Status_B High Resolution / CMPT = 3972
NoteFor further information please see the IODD PDF file under "Downloads"
Factory setting / CMPT = 2
Profiles
Smart Sensor - SSP 0Generic Profiled Sensor
FunctionDevice identification
FunctionProcess data variable
FunctionDevice diagnosis
Min. process cycle time[ms]2.3
IO-Link resolution pressure[bar]0.1
IO-Link process data (cyclical)
functionbit length
pressure14
binary switching information2
IO-Link functions (acyclical)application specific tag
Status_B High Resolution / CMPT = 3
Profiles
Smart Sensor - SSP 3.1Measuring Sensor
Common - I&DIdentification and Diagnosis
Min. process cycle time[ms]3
IO-Link resolution pressure[bar]0.05
IO-Link process data (cyclical)
functionbit length
pressure16
device status4
binary switching information2
IO-Link functions (acyclical)application specific tag
Operating conditions
Ambient temperature[°C]-25...80
Storage temperature[°C]-40...100
ProtectionIP 65; IP 67
Tests / approvals
EMC
DIN EN 61000-6-2
DIN EN 61000-6-3
Shock resistance
DIN EN 60068-2-2750 g (11 ms)
Vibration resistance
DIN EN 60068-2-620 g (10...2000 Hz)
MTTF[years]138
UL approval
UL approval no.J013
Pressure Equipment DirectiveSound engineering practice; can be used for group 2 fluids; group 1 fluids on request
Mechanical data
Weight[g]300
Housingcylindrical
Dimensions[mm]Ø 34 / L = 92.7
Materialsstainless steel (316L/1.4404); PBT+PC-GF30; PBT-GF20; PC
Materials (wetted parts)stainless steel (316L/1.4404); Al2O3 (ceramics); FKM
Min. pressure cycles100 million
Tightening torque[Nm]25...35; (recommended tightening torque; depends on the lubrication, the seal and the pressure load)
Process connectionthreaded connection G 1/4 external thread (DIN EN ISO 1179-2); internal thread:M5
Process connection sealingFKM (DIN EN ISO 1179-2)
Restrictor element integratedno (can be retrofitted)
Displays / operating elements
Display
Display unit3 x LED, green (bar, psi, MPa)
switching status2 x LED, yellow
measured valuesalphanumeric display, red/green 4-digit
Remarks
Pack quantity1 pcs.
Electrical connection
Connector1 x M12; coding: A; Contacts: 4, gold-plated

下载资料